Statement of Work   

 

1.0     Introduction   

 

1.1 This  Statement  of  Work  describes  the  requirements  for  “Design  and  Fabrication  of 

calibration accessories and calibration of KARI wind tunnel balances”.   

 

2.0 

KARI wind tunnel balances   

 

The Supplier shall provide calibration reports and calibration accessories for 3 KARI balances. 

 

2.1     Balance Configurations   

 

ID 

Diameter 

Normal 

Force 

Pitching 

Moment 

Side 

Force 

Yawing 

Moment 

Rolling 

Moment 

Axial 

Force 

 

Inches 

Lbs 

In-Lbs 

Lbs 

In-Lbs 

In-Lbs 

Lbs 

IB6-017 

2.0 

1000 

4000 

500 

2500 

2500 

250 

IB6-018 

2.0 

2000 

6000 

1000 

3000 

2500 

250 

IB6-019 

1.25 

800 

2000 

400 

1100 

600 

125 

   

 

IB6-017 

 


background image

 

IB6-018 

 

IB6-019 

IB6-017 has 7 full bridges, 6-component forces and moments and one additional axial force. 

IB6-018 has 8 full bridges, 6-component forces and moments and two additional axial force. 

IB6-019 has 6 full bridges, 6-component forces and moments. 

 

3.0 

Balance Calibration 

 

3.1      Calibration loading schedule   

 

Primary  load  for  each  axis  and  two-combined  load  will  be  applied  to  the  balance.  Maximum 

loading points for each balances will be about 3200 points. Details about the loading points will 

be decided with supplier’s suggestion. KARI’s suggestion is attached as a appendix. 

 

3.1.1  Primary  loading  :  Primary  load  for  each  components  will  be  applied  to  the  balance  with 

more than 5 steps from zero to a full load condition. 

 


background image

3.1.2 Combined loading ; Two-combined load for all components will be applied to the balance 

to achieve the 6x97 calibration matrix. 

 

3.1.3 The balance shall be designed so that failure (other than a taper failure) during a test does 

not result in the model separating from the sting.   

 

3.1.4    The recommended excitation voltage is 5V.   

 

3.2      Calibration accessories 

 

Supplier  shall  design  and  fabricate  all  accessories  including  the  calibration  body,  balance 

holders and free end adapter for all three balances. Taper gages will be provide to Supplier by 

KARI. 

 

4.0     Schedule   

 

All calibration shall be finished within 8 months after contraction.   

 

5.0     Deliverables   

 

Shipment and insurance for both side is responsibility of KARI. 

A.  Return 3 balances and taper gages. 

B.  Calibration report with electric file 

  The test results from all loadings shall be included, as well as the deflection data. 

The constants, equations and standard deviations will also be included. 

C.  Calibration accessories 

  Calibration Accessories made by this project 

  Electric drawing file for calibration accessories 

 

6.0     Qualification of Supplier 

 

Supplier shall be able to calibrate the KARI wind tunnel balances by designing and fabricate all 

accessories  including  the  calibration  body,  balance  holders  and  free  end  adapter  described  in 

this Statement of Work.   

 

Supplier  shall  provide  the  evidence  of  the  appropriate  equipment  and  facility  for  calibration  of 

the wind tunnel balances described in the Statement of Work. 

 


background image

Appendix 1 : Loading schedule (KARI’s suggestion)   

 

Loading scheme 

 

 

V1 = {0,-50,-100,-75,-50,-40,-20,-10, 0, 10, 20, 40, 50, 75, 100, 60, 30, 0} % of Maximum load range 

V2 = {-100, -75, -60, -45, -30, -15, 15, 30, 45, 60, 75, 100} % of Maximum load range 

 

* Load Rhombus should be applied on {Fy1 + Fy2} and {Fz1 + Fz2} loads 

 

 

 

 

 

 

 

 

 

 


background image

Appendix 2 : Reference drawing   

 

 

IB6-017 Taper drawing 

 

 

IB6-018 Taper drawing 

 

 


background image

 

IB6-019 Taper drawing 

 

 

 

 

Example of Calibration accessories