C o m m o d ity D e s c r ip tio n
Item
Cryogenic Liquid Nitrogen Level Transmitter
Description
1. Instrument – 3 ea
- Mounting: Integral with the probe below
- Display: 2 lines by 8 characters or more
- Input Voltage: 12 Vdc ~ 30 Vdc
- Measurement Signal Output: 4 ~ 20 mA
- Explosion proof enclosure
- Calibration Length: 37 cm
- Standard type
2. Probe/Sensor – 3 ea
- Diameter: OD3/8 (3/8“)
- Fitting: MPT (compatible with above instrument
- Active Length: 37 cm
- Overall Length: 70 cm
- Liquid: LN2
- Fitting: 1/2“ NPT
- Standard Type
* Common Specifications
- Working Fluid: Liquid Nitrogen (LN2)
- Fluid Temperature: around –196 ℃
- Accuracy: 0.1 % or better
- Repeatability: 0.05 % or better
- Linearity: 0.1 % or better